Biotech Systec ZHCR (Zero Hysteresis / Constant Run) Degassing Dual Stage (DST) Vacuum Pump Control System with PCB, Analytical, Bottom Mounted, for DEGASi Degassers - Includes Pump, PCB and Air Bleed
Part Number: 9000-1920
This Degassing Vacuum Control System is designed for simple plug and play operation into most of our degassing chambers. Simply provide 24DC and the system will automatically provide vacuum to the specified vacuum level. Built in continuous head purges ensure that moisture and vapor are swept out of the pump to maintain consistent performance. An optional connector is available for LED error outputs.
This all new generation of pumps is characterized of being almost entirely silent – a remarkable upgrade from older versions. It’s also lighter and operates with less vibrations.
The DST Pump with PCB is a vacuum control system for use with degassing chambers. The system includes a dual stage (DST) vacuum pump and a Smart board controller to maintain vacuum levels. There is a connector for optional external LED indicators or error outputs. The prep versions are designed to handle high vapor loads from high flow rate chambers or fluids with low vapor pressure that will cause increased solvent pervaporation through the membrane.
Systec ZHCR Dual Stage (DST) Vacuum Pump Control System Range:
Part Number | Description | Mounting | Scale |
Vacuum Degassing Control System - includes Pump, PCB and Air Bleed | |||
9000-1920 | Vacuum Control System | Bottom | Analytical |
9000-1921 | Vacuum Control System | Side | Analytical |
9000-1922 | Vacuum Control System | Bottom | Prep |
9000-1923 | Vacuum Control System | Side | Prep |
Pump Only | |||
9000-1892 | Vacuum Pump Only | Bottom | Analytical |
9000-1893 (Replaced by 9000-2061) | Vacuum Pump Only | Side | Analytical |
9000-1896 | Vacuum Pump Only | Bottom | Prep |
9000-1897 | Vacuum Pump Only | Side | Prep |
PUMP CONTROL SPECIFICATIONS:
Power Requirements | 24 VDC @ 0.75 Amp max. (< 5 Watts consumption average) | |
Temperature | 50 o C or lower when run in ambient conditions (20- 25 o C) | |
Vacuum Sensor Calibration Accuracy | Prep 80 mmHg +/- 5 mmHg Analytical 50 mmHg +/- 5 mmHg | |
Closed-Loop Control Setpoint | 50 mm Hg absolute (analytical) or 80mm Hg (prep) pump runs at high RPM until near setpoint, then speed is varied to maintain a value of setpoint— load independent. | |
Errors Detected | 1 – Pumpdown: | Prep — Unable to reach 82 mm Hg in 10 minutes. Analytical — Unable to reach 52mm Hg in 10 minutes. |
2 – High Vacuum: | Prep — Vacuum > 110 mmHg for more than 6 min. in the running state (after pumpdown). Analytical — Vacuum > 82 mmHg for more than 6 min. in the running state (after pumpdown). | |
3 – Sensor Signal: | Sensor Failure Detected Sensor signal >800 mm Hg or sensor signal <10 mm Hg. | |
ANALYTICAL VACUUM PUMP TECHNICAL DATA | ||
Air Flow (no vacuum): | 650 SCCM @ 400 RPM; 100 SCCM @ 60 RPM | |
Typical Vacuum Performance: | 50 mmHg @ (60±20) RPM (3 SCCM air flow, closed loop) | |
Pump-down Time: | <1 minute (with 4 MINI degassing channels, 50 cc total internal volume) | |
Pump Head Continuous Purge Air Flow Rate | ~12 SCCM | |
Vacuum Flow Path Materials: | 303 Stainless Steel, Polypropylene, PTFE, EPDM Rubber | |
Expected Lifetime: | :>5 years (continuous run @ 60 RPM) | |
PREP VACUUM PUMP TECHNICAL DATA | ||
Air Flow (no vacuum): | 650 SCCM @ 400 RPM; 100 SCCM @ 60 RPM | |
Typical Vacuum Performance: | 80 mmHg @ (80±20) RPM (3 SCCM air flow, closed loop) | |
Pump-down Time: | <1 minute (with 4 MINI degassing channels, 50 cc total internal volume) | |
Pump Head Continuous Purge Air Flow Rate: | : ~30 SCCM | |
Vacuum Flow Path Materials: | 303 Stainless Steel, Polypropylene, PTFE, EPDM Rubber | |
Expected Lifetime | >5 years (continuous run @ 60 RPM) |
Tags: Degassers, Degassing Spare Parts, Vacuum Control Systems, Vacuum pumps with control board
- Model: 9000-1920
- Manufactured by: Biotech