HT3000A Series Liquid Gas Chromatography Autosamplers
HT3000A Automatic Liquid Sampler for GC, 15 vials, 2ml Capacity
Offering reliability and superior performance for all GC-GC/MS applications, meeting the requirements of both today and tomorrow.
• 50+ GC models supported • 1 turret serves 2 injectors • Extraordinary performance • Touch screen powered • Move freely between GCs
Analysing fewer samples? Get precision and performance with a system that is perfectly sized for your needs. The HT3100A offers automation for processing up to 15 unattended samples. Quick sets of autosampler parameters can be done by the free PC control (HTA Autosampler Manager), provided together with the HT3100A, while all the routine operations such as analysis start, sample loading and extra wash can be managed by its dedicated keypad.
KEY FEATURES
Easy to use
Just load the samples and run the analysis with no extra downtime. The new run-all–samples mode makes the operation of sample processing even more simple and transparent. HTA is the first to offer GC liquid autosamplers with a large, full-color touch screen interface, providing easier system accessibility and usability. The touch screen eliminates drilldown, simplifying instrument control for both novices and experienced users.
Leader in compactness
The HT3000A Series is the most compact autosampler on the market, with a nearto-zero requirement for bench space, while offering a top-class sample capacity. The HT3000A Series sample rack can be removed for sample loading or preparation, or for storage elsewhere.
Support all GCs of your lab. Perfect for GC/MS
The system is fully self-contained and can be easily installed on all the GC and GC/MS systems available in the market, because of flexibility and modular configuration. It can be configured to serve up to two injectors in most supported GCs. The rotating tower leaves the injector port free for manual injection or maintenance. In addition to avoiding any complicated sample vial transfer procedures, this approach keeps the samples away from any heated source. In fact, the sample rack is mounted away from the GC oven to prevent exposure to high temperatures, which could cause degradation or condensation in the sample vial. The self-aligning “plug and play” HT3000A Series mounts in seconds, without tools. It can be easily moved between GCs when workloads change, due to its easy positioning, repositioning and easy removal.
The HT3000A Series are the fastest liquid autosamplers available on the market: injection is performed in less than 100ms. Fast-injection technology ensures the best possible peak shape, while maximising the accuracy of your results. The greater solvent capacity means longer unattended operations. The autosampler supports the use of six solvent vials of 10ml each, giving a total capacity of 60ml. Furthermore, advantage can be taken of the double wash step capability: as preand post-wash solvent in addition to A, B… F solvents you can also choose for a combination of A+B, A+C…. F+E solvents for superior analytical performance (carry-over adverse!).
HT3000A Series handles the most sophisticated sampling techniques:
internal standard technique (also know as sandwich injection), multi-phase, ambient headspace, priority injection, nanoliter injection and much more. Parameters are easily programmable to optimise both the most convenient sampling methods for both extremely volatile or viscous samples and the best injection technique. Variable needle depths allow you to samples to be taken from anywhere within the vial, thus performing extractions directly from the vial. From small-volume injection to largevolume injection and multi-phase sampling, HT3000A Series can help to process samples more quickly and to access better data. It can also mount different types of syringes, with volumes from 0.5 to 100ml. HT3000A Series offers syringe illumination to always keep the sample under control, in order to check against air bubbles in method validations.
COMPARE MODELS
HTA 3000A SERIES LIQUID GAS CHROMATOGRAPHY AUTOSAMPLERS
Model: | HT3100A | HT3000A | HT3200A |
Sample Capacity | 15 samples | 121 samples (1 removable rack) |
209 samples (2 removable racks) |
User Interface | Keypad | Touch Screen | Touch Screen |
Syringe Housing Illumination | - | √ | √ |
SyringeID2 | - | - | √ |
BCR (Bar Code Reader)1 | - | - | Optional |
Software: HTA Autosampler Manager (Standard Version) | Included | Free trial (60 days) | Free trial (60 days) |
Easy fixing/ Repositioning system | √ | √ | √ |
TECHNICAL SPECIFICATIONS
General features | |
Syringe volume4: | 0.5, 1, 5, 10, 25, 50 and 100 μL |
Maintenance: | preventive maintenance capabilities |
Electrical control: | LAN and TTL; optional: RS232 |
Syringe area illumination: | yes (programmable) for 3000A/3200A |
Tray capacity4 | |
3000A: | 121 vials, 2ml (1 removable rack) |
3200A: | 209 vials, 2ml (2 removable racks) |
3100A: | 15 vials, 2ml |
Filling | |
Sample volume: | as low as step of 0.1μL |
Air volume: | as low as step of 0.1μL |
Filling speed: | 1-100μL/sec |
Viscosity delay: | 0-15s |
Bubble elimination: | up to 15 pull up strokes |
Injection | |
Injection speed: | 1-100μL/sec |
Injection depth: | programmable |
Pre and post inj delay: | 0-99s |
Washing | |
Type: | pre-injection, sample, post-injection |
Solvent capacity: | 6x 10ml vials |
Mode: | single or double wash |
Internal standard technique | |
IS volume: | as low as step of 0.1μL |
Air gap volume: | as low as step of 0.1μL |
Mode: | 1 or 2 air gaps |
Physical features | |
Dimensions (WxHxD)5: | 280 x 570 x 320 mm |
Weight (3000A): | 8.0kg |
Weight (3200A): | 9.2kg |
Weight (3100A): | 6.4kg |
Power supply: | 100-240±10%Vac; 50-60Hz; 55W |
Software | |
HTA Monitor: | included for free |
Virtual Touch Screen and AI functionalities | |
HTA Autosampler Manager: | 60-days free trial Full autosampler PC programming |
HTA Monitor PC requirements | |
Software: |
|
Hardware: |
|
5 minutes to move the autosampler from one GC to another