Instrument:
Thermo Fisher: Helios & Aquamate range
Electrical:
• Filament / Heater Voltage:
- Start: 10 VDC
- Operating: 5 - 7 VDC
• Filament / Heater Current:
- 1.5A Max
• Discharge / Arc Voltage:
- Strike: @25°C >350 VDC
- Operating: 60 - 90 VDC
• Discharge / Arc Current:
- 10 -300mA
• Leads:
- Black: Filament & Cathode
- Red: Anode
• Terminals:
- 3 Pin housing
Optical:
• Window material:
- UV Transmitting Borosilicate
• Source aperture:
- Ø1mm Diameter
• Wavelength range:
- 185nm - 400nm
• Noise:
- <1 x 10-4 Absorption units
• Drift:
- < 3 x 10-3 Absorption units
Operational Life:
When radiant energy reaches 50% of the initial energy at 300mA(dc).
Thermo Fisher: Helios & Aquamate range
Electrical:
• Filament / Heater Voltage:
- Start: 10 VDC
- Operating: 5 - 7 VDC
• Filament / Heater Current:
- 1.5A Max
• Discharge / Arc Voltage:
- Strike: @25°C >350 VDC
- Operating: 60 - 90 VDC
• Discharge / Arc Current:
- 10 -300mA
• Leads:
- Black: Filament & Cathode
- Red: Anode
• Terminals:
- 3 Pin housing
Optical:
• Window material:
- UV Transmitting Borosilicate
• Source aperture:
- Ø1mm Diameter
• Wavelength range:
- 185nm - 400nm
• Noise:
- <1 x 10-4 Absorption units
• Drift:
- < 3 x 10-3 Absorption units
Operational Life:
When radiant energy reaches 50% of the initial energy at 300mA(dc).
- Model: P736
- Manufactured by: Photron