HTA HT4000L Lab-Automation Ready LC Autosampler, 42 vials, 6/10/20ml

Price:
Live chat below or contact us for latest price.


Download Data Sheet

HT4000L: The LC Autosampler Lab-Automation Ready

HTA HT4000L Lab-Automation Ready LC Autosampler, 42 vials, 6/10/20ml

For the same price of a conventional HPLC autosampler, you just get more; be ready to review your expectation about HPLC autosamplers! Enjoy the large sample capacity, the support for a wide range of sample vessels, the application versatility, the lab automation capabilities and the possibility to implement new modules and functionalities directly in your lab. Fits any brand HPLC and LC/MS.

KEY FEATURES

Large Sample Capacity and Wide Range of Sample Vessels
HT4000L can handle up to 176 2ml-vials per run or it can work with different kinds of sample vessels. Inject samples out of vials (2/4/6/10/20/40 ml), test tubes, tube blocks or plates including Vacutainer®, Eppendorf® tubes, and Well Format Tubes. Choose from standard racks or enquire for a custom design: custom racks pay for themselves in a few weeks. Furthermore, racks are easily interchangeable in order to allow quick switching across multiple configurations.

Application Versatility
HT4000L fits all liquid chromatography systems. It is able to look after the widest range of applications including analytical, prep, UHPLC, IC, nano, LC/MS… and many more!

Lab Automation Ready
HT4000L can work as a standard HPLC autosampler. However, you may want to delegate some sample preparation tasks: HT4000L can take care of them as no standard autosampler has been able to assist. Dilution, pre-column derivatisation, pH spiking and internal standard addition are available within the standard version. If standard sample preparation capabilities are not enough, get the HTAPREP option by which you can add support to complex and custom autosampler methods: endless possibilities, any sample preparation procedure or complex sample introduction become available.

Modular and Scalable
A wide range of accessories can be sourced and added in the field at any time; they are available to implement your autosampler with. They include: cooling module, vortex, vial heater and reaction cell, barcode reader (BCR), internal/external needle washing, flow cell, Mini-Uniprep® filter vials and more. We provide your LC autosampler at the best possible price; only at the time your needs change you will pay for the optional accessories to perform the new tasks without the hassle of buying a new autosampler. Your investment is future proof!

Enjoy your daily experience with HT4000L
Easy to use, easy to install and easy to service. We believe in a “robust by design” approach: HT4000L – based on cutting edge technologies and DFX design – well incarnates this philosophy making itself the ideal instrument for installation in any scenario. Set your method once using the complimentary HTA Autosampler Manager, then you just need to load samples and press START: experience a seamless integration with your Chromatography Data System.

TECHNICAL SPECIFICATIONS

General features
Syringe volume:Standard: 100, 250, 500ml and 1, 2.5, 10ml
Nano-Volume Opt.: 5, 10, 25, 100ml
Tray capacity (standard):
  • 176 vials, 2ml (1 removable rack)
  • 96 vials, 4ml (1 removable rack)
  • 54 vials, 6/10/20ml (1 removable rack)
  • 3 96-well plates (1 removable rack)
  • Custom trays available
Tray capacity (cooled):
  • 121 vials, 2ml
  • 42 vials, 6/10/20ml
  • 2 96-well plates
Maintenance:Preventive counters available
Electrical control:LAN and TTL
Filling
Sample volume:as low as step of 1ml3
Air volume:as low as step of 1ml3
Filling speed:1-100ml/sec3
Viscosity delay:0-25s
Bubble elimination:up to 15 pull up strokes
Analytical performanceInjection
RSD:Full loop: ≤ 0.25% (typically 0.15%)
Partial loop: ≤ 0.5%
Injection mode:Full Loop, Partial Loop
Linearity (r):≥ 0.9999Flush Volume:programmable
Carry over2:≤ 0.05% (with standard wash)
typically ≤ 0.01% (with int/ext needle wash)
Injection speed:1-500ml/sec3
Wash (Pre & Post Injection)Analysis time:0-999min
Washing cycle:up to 15 (Mode: every injection/sample/step)Wait time between consecutive injections:as low as 12sec4
Reagents:up to 2 different reagents for wash functionPhysical features
Stator wash:programmableDimensions (WxHxD):310x705x690mm
Dilution, derivatization, pH spiking, IS addictionWeight:15kg
Sample dispatch:step of 1ml3Power supply:100-240±10%Vac; 50-60Hz; 55W
Reagent addiction:step of 1ml3; up to 3 different reagentsCooling Module
Mixing cycle:programmableTechnology:Peltier
Intermediate wash cycle:up to 15Temperature range conditioning:from 4°C to 40°C
Reaction time:0-100 minutes (0.1 minutes step)Temperature accuracy:+/-1°C (typically +/-0.5°C) between the vials
Additional modules:
Among the most popular modules:
Cooling capacity:about 20°C below the RT (on the vial)
  • Barcode reader (BCR)
  • Flow Cell
  • Gripper for vial displacement
  • Internal/external needle washing
  • Sample Capacity Extension
  • Vial Heater/ Reaction Cell
  • Vortex
Working ambient temperature:25°C +/-10°C
Weight:about 5kg
Power supply:100-240±10%Vac; 50-60Hz; 120W
Field installation:yes

1 Tested with Uracile 100ppm in mobile phase 60% methanol : 40% water; Column C18; detector: UV fixed length; loop: 20ml
2 Without accounting the transfer line

HT4000L Operations

 


  • Model: HT4000L
  • Manufactured by: HTA