SONIMIX 7100 MFC Gas Mixer
Available in the Following Configurations:
Sonimix 7100-2L1R | 2 Lines 1 Range gas blender, Dilution range 10% to 100%, Total flow 3.5 Nl/min |
2 Lines 1 Range gas blender, Dilution range 1% to 10%, Total flow 3.5 Nl/min | |
2 Lines 1 Range gas blender, Dilution range 0.1 to 1%, Total flow 3.5 Nl/min | |
Sonimix 7100-2L2R | 2 Lines 2 Range gas blender, Dilution range 1% to 100%, Total flow 3.5 Nl/min |
2 Lines 2 Range gas blender, Dilution range 0.1% to 10%, Total flow 3.5 Nl/min | |
Sonimix 7100-2L3R | 2 Lines 3 Range gas blender, Dilution range 0.1% to 1% + 1% to 10% + 10% to 100%. Total flow 3.5 Nl/min |
Sonimix 7100-3 / Multigas | Gas blending system with 3 MFC/lines and 3 independent inlets, * |
Sonimix 7100-4 / Multigas | Gas blending system with 4 MFC/lines and 4 independent inlets, * |
Sonimix 7100-5 / Multigas | Gas blending system with 5 MFC/lines and 5 independent inlets, * |
Sonimix 7100-6 / Multigas | Gas blending system with 6 MFC/lines and 6 independent inlets, * |
*MFC flow have to be defined at the order
The Sonimix 7100 is a gas mixer mainly used to calibrate and linearize gas monitors and to generate specific gas mixtures.
The Sonimix 7100 is based on the principle of gas flow through MFC (Mass Flow Controllers) and is built according the ISO 6145/7 standards. This instrument provides a very high flexibility allowing to combine multigas mixtures (up to 5 gases + a balance gas) and multi-range functionality.
The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.
All models make automatic calculation in function of the used gases types. The Sonimix 7100 is totally compatible with the US EPA methods and can be easily linked with National Standards (NIST, METAS, UKAS…)
Each mass flow controller is factory calibrated in 10 points by a traceable primary standard.
Product features:
- Fulfilment of ISO 14481 standards CEM (QUAL 1 to QUAL 3)
- ISO 6145/7 standards
- High accuracy - (2% rel.)
- constant gas flow and pressure
- high dilution range - 0.1% to 100%
- No maintenance
- Cost savings – Reduce gas cylinders costs
- Corrosive gases can be used with the optional SilcoNert © coating
- Remote control
- Ergonomic with LCD display
Main applications:
- Quality control of gas analysers and sensors
- Gas analysers and sensors production
- Interference measurement
- Fulfilment of ISO 14481 standards CEM (QUAL 1 to QUAL 3)
Options:
- 6800 0x x50 - Alternative flow Total flow 5 Nl/min not available for Sx7100-Multigas not available for Sx7100-Multigas included
- X=1 or 3 - Additional ranges 1 line with extended range
- X=2 or 4 - Additional ranges 1 linde double extended or 2 line single extended
- X= 5 - Additional ranges 1 line with extended range and 1 line double extended
- 6800 11 000 - Additional inlets Optionnal 3 additional inlets (4 inlets in total) only for 2LXR and L1 L2 and L3 on Multi-lines
- Silconert 2000 - Silconert 2000 Price by MFC
- 6800 10 000 - Portable casing for casing 19'' Option portative 50x19x40 cm in stead of the 19 Inch rack casing. * Only for multi-range
- 6980 30 010 - ISO 17025 New device Additional cost for ISO 17025 calibration for new Sx7100-2L1R*
- 6980 30 020 - ISO 17025 New device Additional cost for ISO 17025 calibration for new Sx7100-2L2R* 1.
- 6980 30 030 - ISO 17025 New device Additional cost for ISO 17025 calibration for new Sx7100-2L3R* 1.
- 6980 30 040 - ISO 17025 New device Additional cost for ISO 17025 calibration for new Sx7100-Multigas* Per MFC
* UKAS, COFRAC, SIT, SAS compatible
Technical Specifications:
Models: Sonimix 7100 | Sx 7100 2L1R |
Sx 7100 2L2R |
Sx 7100 2L3R |
Sx 7100 Multi Gas |
General data | ||||
Dilution range | 10 – 100 % | 1 – 100 % | 0,1 – 100 % | At configuration, mix up to 6 cylinders |
Dilution principle | Mass Flow Controllers | |||
Dilution gas line flow accuracy | 0.5 % (rel.) + 0,1 FS | |||
Source gas line flow accuracy | 0.5 % (rel.) + 0,1 FS | |||
Repeatability on flow | 0.2% (rel.)+ 0,1 FS | |||
Accuracy on dilution | Better than 2% (rel.) according to ratio dilution | |||
Warm up time | < 45 min | |||
Stabilization time | less than 1 min (once the device has warmed up) | |||
Tightness | < 10-7 mbar l/s verified by He detection | |||
Operating temperature | 15°C - 40°C | |||
Dimensions | 19" 3HE/84 TE , 500mm deep portable casing in option | |||
Net weight | 15 - 20 Kg | |||
Gas inlet requirments | ||||
Dilution gas | Requires dry (-40°C DP) Air, Nitrogen or other carrier gas | |||
Source gas | compatible with low corrosive gases | |||
Pressure | 3 ± 0.3 bar rel | |||
Fittings | Swagelok 1/4" | |||
Gas outlet requirments | ||||
Flow | 3.5 L/min (other flow rate upon request) | |||
Pressure | atm + 150 mb max (other pressure upon request) | |||
Fittings | Swagelok 1/4" | |||
Communication | ||||
RS 232 | Standard | |||
Ethernet TCP/IP | Standard | |||
USB | Standard | |||
Electrical data | ||||
Power supply voltage | 230V/50-60Hz 115V/60Hz 100V/50-60Hz |
|||
Installed power (Max) | 65 - 150W |
- Model: SX 7100
- Manufactured by: LNI SWISSGAS