SONIMIX 7050 MFC Gas mixer
Available in the Following Configurations:
Sonimix 7050-2L1R
SX 7050 2L1R |
2 Lines 1 Range gas blender, Dilution range 10% to 100%, Total flow 3.5 Nl/min |
2 Lines 1 Range gas blender, Dilution range 10% to 100%, Total flow 5 Nl/min |
Sonimix 7050-2L2R
SX 7050 2L2R |
2 Lines 2 Range gas blender, Dilution range 1% to 100%, Total flow 3.5 Nl/min |
2 Lines 2 Range gas blender, Dilution range 1% to 100%, Total flow 5 Nl/min |
Sonimix 7050-2L3R
SX 7050 2L3R |
2 Lines 3 Range gas blender, Dilution range 0.1% to 1% + 1% to 10% + 10% to 100%. Total flow 3.5 Nl/min |
2 Lines 3 Range gas blender, Dilution range 0.1% to 1% + 1% to 10% + 10% to 100%. Total flow 5 Nl/min |
xx = 35 for 3,5 L/min total flow or xx = 50 for 5 L/min total flow
The Sonimix 7050 is a simplified gas mixer based on the principle of gas stream through mass flow controllers (MFC) and mainly used to calibrate and linearize Gas Monitors. Built according to the ISO 6145/7 normalization, the Sonimix 7050 is totally compatible with the US EPA methods and can be easily linked with National Standards (NIST, METAS, UKAS…).
The Sonimix 7050 Multi range gas mixer performs a large range of gas mixtures from high concentration gas cylinders
Product Features:
- Good accuracy and stability
- Remote control
- Maintenance free
- High dilution range
- US EPA method
Main applications:
- To fulfill the ISO 14 481 standard
- Laboratory (quality control of gas Monitors)
- Interference measurement
- Gas monitors / gas sensors manufacturer calibration
- To generate specific atmospheres
Options :
- 6800 10 000 - Portable casing for casing 19'' Option portative 50x19x40 cm in stead of the 19 Inch rack casing
- 6980 30 010 - ISO 17025 - New device Additional cost for ISO 17025 calibration for new Sx7050-2L1R*
- 6980 30 020 - ISO 17025 - New device Additional cost for ISO 17025 calibration for new Sx7050-2L2R*
- 6980 30 030 - ISO 17025 - New device Additional cost for ISO 17025 calibration for new Sx7050-2L3R*
* UKAS, COFRAC, SIT, SAS compatible
Technical specifications:
Model: SX 7050 |
SX 7050 2L1R |
SX 7050 2L2R |
SX 7050 2L3R |
Flow rate (Max) |
3,5 or 5 L /min (other flow under request) |
Dilution |
MFC (mass Flow Controllers) |
Dilution range |
10% to 100% |
1% to 100% |
0.1% to 100% |
Carrier gas |
Requires dry (-40°C DP) Air, Nitrogen or other |
Accuracy of Flow |
±(0,5% RD + 0,1% FS) |
Repeatability of Flow |
0.2% RD |
Flow control Stability |
0.1% FS |
Flow Temperature Sensitivity |
zero: < 0,05% FS/°C; span: < 0,05% Rd/°C |
Flow Pressure sensitivity |
0,1% Rd/bar typical N2; 0,01% Rd/bar typical H2 |
Inlet pressure |
3 bar rel |
Outlet pressure |
atm + 150 mb max (other pressure upon request) |
Warm up time |
< 45 min |
Stabilization time |
< 1 min (once the device has warmed up) |
Communication |
Connections |
Swagelok 1/4" |
Remote command |
RS 232 / Ethernet / USB |
General data |
Supply voltage |
230V/50-60Hz - 115V/60Hz - 100V/50-60Hz |
Rated power (max) |
min 65 to max 150 [Watt] |
Dimensions |
19" 3HE/84 TE , 500mm deep / portable casing in option |
Net weight |
15 to 20 Kg |