THIS PRODUCT HAS BEEN REPLACED BY PRODUCT NUMBER 0001-0220 (SEE LINK BELOW)
0001-0220 - Systec AF Biotech DEGASi Prep+ Preparative Degasser, 2-channel, 100 ml/min per channel
Systec Biotech DEGASi Preparative Stand Alone Degasser, 2-Channel, Prep 20mL/min
- No. of Channels :- 2
- Volume Per Channel :- 8.4
- Max HPLC Gradient Flow CapabilityC (mL/min) :- 20
- Pressure DropD (kPa/mL/min) :- 0.04
The Systec Stand Alone Prep Scale vacuum degassing system for HPLC is a high efficiency in line module that removes dissolved gasses from HPLC mobile phases. Built for the rigors of modern preparative and semi-prep scale HPLC, its unique design assures reliable continuous operation and the highest level of continuous performance available. Two solvent lines may be degassed simultaneously by one unit.
Features
- Ultra-high degassing efficiency
- Low volume, easy to prime
- ZHCR® patented control eliminates baseline fluctuations
- Dual lumen design for low-flow resistance
- Inert flow path
- 5+ year lifetime
- CE & RoHS compliant
- 2-pin vacuum level validation output
- ZHCR® Control with Built-in Test Diagnostics
- Microcontroller self-test vacuum sensor validation on power-up
- Continuous vacuum system monitoring to ensure optimum operational conditions are maintained
- Vacuum system fault detection and shutdown function indicators
- Systec AF / ZHCR Degassing Technology
The prep / semi-prep scale degassing chambers are designed to be easy to prime, and are configured with a Systec AF degassing membrane to provide maximum degassing capacity with the absolute minimum internal volume (<3% of PTFE designs with comparable degassing capacity). The ZHCR (Zero Hysteresis / Constant Run) vacuum pump employs a patented closed-loop, micro-stepping RPM control strategy permitting the pump to run with continuously variable speed, providing quick pull-down at high RPM, and then sustaining a consistent vacuum level at low RPM. Fluctuations in detector baseline due to changes in vacuum level are eliminated by not having to repeatedly stop and start a single-speed pump. This also greatly reduces wear and noise. The brushless motor enables quiet operation and is appropriate for environments where solvent vapors may be present.
Validation Output
A 2-pin rear panel receptacle labeled “Validation” with mating screw-lock plug is provided to allow a validation signal from the control circuit to be sent to a computer or
data system. This validation output indicates vacuum level.
Signal:
5mV DC / 0.13kPa (1 mmHg) absolute from 2.7 to 106.7 kPa (20 to 800 mmHg) (0.100V DC at 2.7 kPa (20 mmHg); 4.000V DC at 106.7 kPa (800 mmHg)
Accuracy:
±1.0% of reading ±0.010V DC from 2.7 to 106.7 kPa (20 to 800 mmHg)
Power Requirement
- Input Power required with AC Adapter (included): 100 to 240V AC (±10%), 1A, 50 to 60 Hz (±3 Hz)
- Four interchangeable wall sockets are supplied with the AC Adapter: North America/Japan, U.K., Continental Europe, Australia.
CE Certification
This product has been certified under the following CE testing standards: EN61326-1; EN55011; EN61300-3-2; EN61300-3-3 & EN61010-1.
Specifications
- Stand-Alone Degassing Modules
- Number of Degassing Channels: 2
- Degassing Efficiency† @ 1 mL / min MeOH > 70% O2 Removal
- Membrane Material: SYSTEC AF
- Other Wetted Materials PEEK, PPS(GF), PTFE(GF), FEP
- Solvent Compatibility: Not compatible with fluorinated solvents.Special version available for GPC solvents.
- Flow Path ID: 1.14 mm (0.045”)
- Internal Volume: 480 μL (standard)
- Maximum Pressure: (@ 25 °C) 0.5 MPa (70 psi)
- Pressure Drop: 0.18 kPa/mL/min
- † Degassing efficiency can be optimized based on flow rate, fluid to be degassed, and gas to be removed. 1 Standard ID; other sizes available.
- Model: 0001-6482
- Manufactured by: Biotech