CTS Sciencix 2 Position/6 Port Rotor Seal, Vespel (600 bar), Agilent 1200, 1260 - CTS-12012, 0101-1416

Price: £403.29


Qty. 
+-
Download Data Sheet

2 Position/6 Port Rotor Seal, Vespel® (600 bar), Comparable to OEM # 0101-1416

Sciencix Part Number: CTS-12012
Equivalent to OEM Part Number: 0101-1416

Compatibility

Agilent® 1200 and 1260 HPLC systems

Operating Specifications

  • Pressure: Max 8700 psi/600 bar
  • Operating Range: 4ºC - 55ºC
  • pH Range: 0-10

Materials Specifications

Vespel®
316 SS

Replacement Frequency

The rotor seal generally is not part of a required maintenance protocol. Indications that the rotor may need to be replaced are poor reproducibility of injection volume or a leak at the injection valve

Operating Recommendations

The typical cause of failure at the rotor seal is wear of the Vespel® material. A major cause of wear is the use of buffered mobile phases. To help improve the lifetime of the rotor seal it is recommended that the system and injection valve be flushed with either the mobile phase without the buffer addition or with an aqueous solution. Be sure to toggle the valve through both the bypass and inject positions to flush all of the mobile phase salts from the injection port.

Installation Information

Sciencix parts are designed to be compatible and interchangeable with OEM parts. Please see instrument service manual for proper removal and installation procedure.


  • Model: CTS-12012
  • Manufactured by: CTS