Manual Injection Valve Rotor Seal, Vespel®, Comparable to OEM # 7125-047 , 09904802 , 0101-0623
Sciencix Part Number: CTS-A019
OEM Part Numbers: 09904802, 0101-0623, 7125-047
Compatibility:
- Agilent® 1100 and 1200
- PerkinElmer® 200 Series
- Rheodyne® 7125, 7126, 7725, 7725i and 9725
Wetted Materials:
- Vespel
- 304L SS
Operating Specifications:
- Pressure: Max 6,000 psi/400 bar
- Operating Range: 4ºC - 80ºC
- pH Range: 0-10
Operating Recommendations
The typical cause of failure at the rotor seal is wear of the Vespel material. A major cause of wear is the use of buffered mobile phases. To help improve the lifetime of the rotor seal it is recommended that the system and injection valve be flushed with either the mobile phase without the buffer addition or with an aqueous solution. Be sure to toggle the valve through both the bypass and inject positions to flush all of the mobile phase salts from the injection port.
Installation Information
Sciencix parts are designed to be compatible and interchangeable with OEM parts. Please see instrument service manual for proper removal and installation procedure.
Replacement Frequency
The rotor seal generally is not part of a required maintenance protocol. Indications that the rotor may need to be replaced are poor reproducibility of injection volume or a leak at the injection valve.
- Model: CTS-A019
- Manufactured by: CTS