CHIRALPAK® ID-U, CHIRALPAK® ID-3 ,CHIRALPAK® ID
These columns use a stationary phase in which an amylose derivative (Amylose tris(3-chlorophenylcarbamate)) is immobilized on a silica gel carrier.
- These mobile phases and sample dissolving solvents, not only use hexane and alcohol, but also ethyl acetate, THF, halogen-based solvents, DMSO, and other solvents, which can be used with silica gel-based HPLC columns.
- It can also be used under reversed phase conditions.
- We have semi-micro (thin diameter), analytical, and semi-preparative columns, so you can select the column size according to your application.
Specifications:
Particle size | CHIRALPAK® ID-U: 1.6μm CHIRALPAK® ID -3: 3μm CHIRALPAK® ID : 5μm |
pressure | CHIRALPAK® ID-U: To extend the life of the column, it is recommended that the pressure does not exceed 70MPa. CHIRALPAK® ID-3 / CHIRALPAK® ID: To extend the life of the column, it is recommended that the pressure does not exceed 30MPa. |
pH | 2.0 to 9.0 |
temperature | 0-40°C (Normal-phase mobile phase) 5-40°C (Reverse-phase mobile phase with pH<7) 5-40°C (Reversed-phase mobile phase with pH?7) |
sealing solvent | n-hexane / 2-propanol = 90 / 10 (v/v) |