Monolithic Capillary Columns
ULTRON HF-ODS (C) C18 Monolithic Capillary Column, Endcapped, 700 mm Length x 0.1 mm ID
Shinwa Chemical Industries Ltd. Part Number: LCC-14107-352
ULTRON HF-ODS
ULTRON HF-ODS is a monolithic silica (hierarchical porous structure) bonded with octadecyl groups inside fused silica capillary tubing. Useful for Proteome analysis (peptide mapping) with high reproducibility and peak capacity.
The outer diameter of the capillary column is 0.38 mm. Please use suitable fittings (Sleeve , Ferrules etc) for your Nano Flow LC system. Normally, FEP sleeve fittings are shipped as a standard specification. Pressure resistant fittings (Inlet side: PEEK sleeve) are also available as option with additional cost.
Precaution for use
- The capillary column is a thin quartz tube and it is delicate towards scratches. Do not subject the column to sudden shock such as striking or dropping, and rubbing with a metal etc. Columns may break due to fine scratches or similar causes.
- The fittings (Pre-fixed) are optimized in advance. Please do not disassemble by removing parts, etc.
- It is possible to bent the column up to R = 100 mm. Be careful not to bend it more strongly as this will cause breakage of the column bed or capillary tube.
- Please use the column at the maximum allowable pressure or less. The maximum allowable pressure is 20.0 MPa for the standard specification (FEP sleeve) and 30.0 MPa for the pressure resistance specification (PEEK sleeve).
- Ensure to filter samples and mobile phases using a membrane filter with a mesh size of 0.45 μm or smaller before using the column. Failure to filter mobile phases, etc. can lead to blockages of column filters and increases in analytical pressures.
- Ensure to thoroughly degas mobile phases prior to use. Insufficient degassing of mobile phases can lead to the formation of bubbles inside analytical instruments and columns resulting in problems with analyses.
- It is recommended to dissolve the sample in a solvent that is of the same composition as the initial mobile phase.
- ULTRON HF-ODS columns are shipped containing 100% methanol. When replacing this solution with mobile phases, in order to avoid the danger of sudden pressure increases, flow the mobile phase at an initial flow rate of 200 nL/min at least 10 minutes (in case of the column length is 250 mm) or at least 30 minutes (in case of column length is 700 mm), and then increase the flow rate to the desired value.
Cleaning and storage
- Please clean the column with a mixture containing organic solvent/water (not including acid and inorganic salt), such as at least 60% organic aqueous solution.
- Do not remove the column from LC system before the pressure drops zero.
- For long-term storage, please replace the column with organic solvent, close the column with end stop plugs tightly and store at cool and dark place.
Specifications
Column name | ULTRON HF-ODS (N) | ULTRON HF-ODS (C) |
---|---|---|
Column type | C18 (non-end capping) | C18 (end capping) |
Column size ( length x inner diameter) | 250 mm x 0.1 mm 700 mm x 0.1 mm |
250 mm x 0.1 mm 700 mm x 0.1 mm |
Applications | ?High resolution peptide mapping and omics | |
Flow rate range | Recommended range: 300?500 nL/min | |
pH range | Recommended range: 2.0?8.0 | |
Recommended temperature range | FEP sleeve: 30?50°C (Max. temp. 50°C) PEEK sleeve: 30?60°C (Max. temp. 60°C) |
|
Max allowable pressure | FEP sleeve: 20.0 MPa PEEK sleeve: 30.0 MPa |
- Model: LCC-14107-352
- Manufactured by: Shinwa Chemical Industries Ltd