BEAM 1.1 (Z5) Laser Gas Generator Open-Loop / Flow Mode
BEAM 1 series : ZERO AIR HC removal < 0.05 ppm
Laser Gas Generator up to 1.7 L/min - Open Loop / Flow mode - Integrated Pump
The BEAM 1.1 performs in 24/7 production process worldwide with utmost reliability. Even under the most challenging environments BEAM 1.1 guarantee frictionless industrial use for different applications (example: high power UV-lasers (>300 nm)). The Generator produces a continuous flow of hydrocarbon-free air. The flow is controlled stable. All Generators reduce the total quantity of hydrocarbons to less than 0,01 ppm, establishing a stable flow for your application. Zero-air produced in this way replaces existing stocks of synthetic air stored in high pressure bottles.
Laser Gas Generators are designed specifically to supply clean, dry, hydrocarbon-free air to be used for several applications.
LNI Swissgas recognizes the importance of having a safe, reliable and cost-effective supply of high-purity Zero Air.
In addition, providing flexibility to add or remove modules as your application requirements evolve over time.
Product features:
- Easy use and quick installation
- RS 485 as standard
- Purity: < 0.05 ppm THC
- Easy maintenance
- Reduced foot print
- Stackable with other LNI series
- integrated pump
Main applications:
PROTECTION FOR LASER
LNI Swissgas generators continually protect optics, crystals and coatings of the laser beam path against particles, humidity, hydrocarbons and aggressive gases. Degradation and maintenance are reduced to a minimum. Lifetime and performance of the laser and optical systems are maximized.
Available Consumables:
Product Code | Periodicity | Description |
8100.10.005 | 4000h | Kit filters for BEAM 1.1 |
8100.10.008 | 8000h | N°3 refill for dessicant column for BEAM 1.0 - 1.1 |
Technical Specifications:
Model: Beam 1.1 | |
Zero Air Outlet | |
Flow rate (Max) | 1.0 l/min |
Outlet pressure (min) | / |
Outlet pressure(max) | 1.2 bars |
Total hydrocarbon content | < 0.05ppm |
Start-up time | 50 min |
Air inlet | |
Max inlet hydrocarbon content | Integrated Pump |
Min supply pressure | |
Max supply pressure | |
Dew Point | |
Temperature (min;max) | |
Communication | |
RS232 | Standard |
General Data | |
Supply voltage | 100-240Vac 50/60Hz |
Connection type | IEC320-C13 |
Installation power (max) | 240W (280VA) |
Fuse rating (5x20mm) | 2A |
Dimensions | Standard 19" - 4U |
Net weight | <13 kg |
Connections | |
Outlet port | 6 mm Push-Fit |
Inlet port | 6 mm Push-Fit |
Health connectot | max 5A, 40VDC MAX / 24VAC MAX (normaly closed) |
Operating conditions | |
Temperature (min;max) | 5-35°C (41-95°F) |
Humidity (max, non condensing) | 80% at 25°C (77°F) |
Noise | <25dB (A) |
Ip rating | IP20 |
- Model: BEAM 1.1 (Z5)
- Manufactured by: LNI SWISSGAS