CTS Sciencix Rotor Seal, Vespel, 2-Groove, 7750-040, Agilent 1100, 1200 - CTS-10459, 0100-1853

Price: £72.71


Qty. 
+-
Download Data Sheet

Rotor Seal, Vespel®, 2-Groove, 7750-040, Comparable to OEM # 0100-1853

Sciencix Part Number: CTS-10459
Equivalent to OEM Part Numbers: 0100-1853

Compatibility

Agilent® 1100 and 1200 HPLC systems

Wetted Materials Specifications

Vespel®
304L SS

Operating Specifications

Pressure: Max 6,000 psi/400 bar
Operating Range: 4ºC - 80ºC
pH Range: 0-10

Replacement Frequency

The rotor seal generally is not part of a required maintenance protocol. Indications that the rotor may need to be replaced are poor reproducibility of injection volume or a leak at the injection valve.

Operating Recommendations

The typical cause of failure at the rotor seal is wear of the Vespel® material. A major cause of wear is the use of buffered mobile phases. To help improve the lifetime of the rotor seal it is recommended that the system and injection valve be flushed with either the mobile phase without the buffer addition or with an aqueous solution. Be sure to toggle the valve through both the bypass and inject positions to flush all of the mobile phase salts from the injection port.

Installation Information

Sciencix parts are designed to be compatible and interchangeable with OEM parts. Please see instrument service manual for proper removal and installation procedure.