Rotor Seal, Vespel®, for p/n 0101-0921, Comparable to OEM # 0100-1855

Sciencix Part Number: CTS-11419
Equivalent to OEM Part Number: 0100-1855

Compatibility:

Agilent® 1100 and 1200 HPLC systems

Operating Specifications:

  • Pressure: Max 6,000 psi/400 bar
  • Operating Range: 4ºC - 80ºC
  • pH Range: 0-10

Wetted Materials:

  • Vespel®
  • 316 SS

Operating Recommendations

The typical cause of failure at the rotor seal is wear of the Vespel® material. A major cause of wear is the use of buffered mobile phases. To help improve the lifetime of the rotor seal it is recommended that the system and injection valve be flushed with either the mobile phase without the buffer addition or with an aqueous solution. Be sure to toggle the valve through both the bypassand inject positions to flush all of the mobile phase salts from the injection port.

Replacement Frequency

The rotor seal generally is not part of a required maintenance protocol. Indications that the rotor may need to be replaced are poor reproducibility of injection volume or a leak at the injection valve.

Installation Information

Sciencix parts are designed to be compatible and interchangeable with OEM parts. Please see instrument service manual for proper removal and installation procedure.