Ceramic Plunger Assembly, Comparable to OEM # 228-35009-92
Sciencix Part Number: CTS-10584
Equivalent to OEM Part Number: 228-35009-92
Compatibility
Shimadzu® LC-10ATvp pumps
Wetted Materials Specifications
316 SS
Ceramic
Replacement Frequency
Replacement may be indicated by a leak at the pump head. The plunger should be inspected with a magnifier whenever the pump head is disassembled to replace the plunger seals. If any scratches are observed or a worn or flat area is seen the plunger should be replaced.
Installation Information
Sciencix parts are designed to be compatible and interchangeable with OEM parts. Please see instrument service manual for proper removal and installation procedure.
- Model: CTS-10584
- Manufactured by: CTS