Ceramic Plunger Assembly, Comparable to OEM # 228-35601-91, Old# 228-34498-92
Sciencix Part Number: CTS-A138
Equivalent to OEM Part Numbers: 228-34498-92 - replaces 228-34498-92
Compatibility
Shimadzu LC-10ADvp, LC-20AD/AB and LC-20ADvp pumps
Materials Specifications
Ceramic
316 SS
Replacement Frequency
Replacement may be indicated by a leak at the pump head. The plunger should be inspected with a magnifier whenever the pump head is disassembled to replace the plunger seals. If any scratches, a worn area or flat area are seen then the plunger should be replaced.
Installation Information
Sciencix parts are designed to be compatible and interchangeable with OEM parts. Please see instrument service manual for proper removal and installation procedure.
- Model: CTS-A138
- Manufactured by: CTS