JETSTREAM II PLUS
The peltier heating/cooling column thermostat with the big compartment room can accommodate up to 4 columns with a length of up to 40 cms, ample space for valves and injectors is provided.
Input is simplest via the alphanumerical keypad or Rs232- Interface. Temperature control completely programmable, isothermal steps / linear gradients. Precise temperature control enhances separation and reproducibility and improves the quality of your analysis.
Using the peltier-technology offers you the possibility of heating and cooling, providing you with complete independence of the room temperature. This feature doesn’t only “come in handy” when working in low temperature ranges; with works having to be done at
room temperature, a cooling might be necessary to grant exact steering.
Double-sensor-reference-technology and auto-calibration.
Recurrent precise analysis... ...guaranteed
Peltier heating / cooling column thermostat,fully programmable; isothermal steps and linear
gradients, for 4 columns with a length of up to 40cm, integrated leaksensor with adjustable
sensitivity,column-protection with temperature lock-feature valves and injectors can be built in.
- large Temperature-range (5°-85° C)
- highest stability (+/- 0.02° C) !!
- and reproducibility (+/- 0.1° C)
- by the DSR (Double-Sensor-Reference-technology)
- Auto-calibration and additional correction via keyboard,
- Leak sensor with adjustable sensitivity, auto-switch-off
- and "T-Lock"- column protection feature
- highest level of security through continous function monitoring
- Rs232-Interface for GLP-Conformity
Specifications
Temp. Range 5-85°
Temp. Accuracy +/- 0,1 °
Temp. Stability +/- 0.02 °
Power supply 100-245VAC 50/60Hz 100W
Outline dimensions 135(W)x310(L)x450(H) mm,
Weight 11 k
Heating / Cooling System
The 2-way-forced-air-circulation guarantees perfectly even distribution of the thermostated air
within the column compartment.
+ Temperature control by Double-Sensor-Reference-Technology:
2 independent sensors, one laser trimmed, the other one, a PT as it is used in
most of the precise instruments in continuous comparison to a precision-reference.
Column Compartment
Thermo-insulated from the environment, accommodate up to 4 columns with a length of up to 40 cm.
Ample space provided for pre-columns injector-valve and column-switch valve.
Usable in horizontal and vertical position.
+ compact outline dimensions for small installation space
Electronics / Programming / External Control
Self check and autocalibration at power-on, self-explaining instructions by 2-line liquid crystal display
linear ramp calculations or steps can be programmed. Selectable parameters are: temperature, duration
and rate of heating/cooling.
+ Input of temperature in Celsius or Fahrenheit,
+ Rs232 Windows application, control over voltage free contact
Autocalibration
To ease in-house-quality-controls, we have created the auto-calibration-feature. After the input of a certain
combination over the keypad, the column-thermostat will start an 8-hour-self-calibration- and control-routine.
In addition, temperatures can be corrected via the keypad.
+ Auto-Calibration-Feature and temperature-correction easiest via keypad
Security
Integrated leak detection with adjustable sensitivity, acoustic warning signal and auto-switch-off.
Temperature-Security-Feature (T-Lock) setable overtemperature protection to prevent destruction o
f temperature sensitive columns caused by handling errors.
Continuous function monitoring and overtemperature protection for general safety.
+ 2 independent sensors and Fuzzy Control Software for controlling
- Model: 5480
- Manufactured by: NON Branded Products