ULBON HR-SS-10 Polyimide Coated Fused Silica GC Capillary Columns
HR-SS-10 - Nitrile silicone
“ULBON” is a generic name of the Shinwa capillary columns.
“ULBON HR” is a generic name of the Shinwa capillary columns with chemical bonding layer.
The capillary column performs very higher separation than the packed column because of the small
diffusion and good permeability of the sample. The fused silica capillary column coated polyimide resin on the outside of tubing was developed in around 1980. The inert surface and flexibility of fused silica tubing lead to high performance separation and easy handling.
ULBON HR Series
Shinwa has ten types of stationary phases for general analysis and each analytical purpose, as an ULBON HR series. They also stock the columns with various lengths, inner diameter and the film thickness of stationary phase for your requirement.
Characteristics
The inert inner surface of the column results in high quantitative analysis.
The adsorption point of the inner surface of the column has been completely removed by a rigid
treatment. Therefore, free acid and free amine which are easily adsorbed are eluted without interaction with the active surface.
No exfoliation and bleeding of stationary phase lead to high sensitive analysis.
In HR series column, the rigid chemical bonding is treated on the surface, therefore there is almost no exfoliation and bleeding of stationary phase. No bleeding from the column results in the high sensitive analysis.
The maximum temperature is higher than that of ordinary WCOT column.
The maximum temperature of HR-20M (correspondence to PEG-20M) is 240 °C and HR-silicone
series are 330 °C (except HR-1701 and HR-17). The maximum temperature of chemical bond type
column rises about 70 °C compared with the WCOT type.
ULBON HR Series (General purpose type)
Column | Stationary Phase | Polarity |
Maximum Temperature |
Usage |
HR-1 (Corresponding OV-1, OV-101) |
Dimethyl silicone resin | None | 330 °C | General analysis |
HR-52 (Corresponding SE-52) | 5% Phenylmethyl silicone | Low | 330 °C | General analysis |
HR-17 (Corresponding OV-17) | 50% Phenylmethyl silicone | Mid | 300 °C | Medicine, saccharide, TMS derivatives, phenols, steroids, and solvents |
HR-1701 (Corresponding OV-1701) | 14 % Cyanopropyl phenylmethyl silicone | 280 °C | ||
HR-20M (Corresponding PEG-20M) | Polyethyleneglycol | High | 240 °C | General analysis, polar compounds, esters, ketones and alcohols |
ULBON HR Series (Special purpose type)
Column | Stationary Phase | Polarity | Maximum Temperature | Usage |
HR-SS-10 | Nitrile silicone | High | 220 °C / 230 °C | Fatty acid ester (especially, geometric isomer of C1=18 and C2=18) |
HR-Thermon-3000B | Phthalic acid ester | High | 220 °C / 230 °C | Fatty acid esters (analysis of unsaturated C20) |
HR-Thermon-600T | Phthalic ester | High | 240 °C | Essential oil, Perfume |
HR-Thermon-HG | Alkyleneglycol Phthalic acid ester polymer | High | 160 °C | Methylmercury, Ethyl mercury, Toribtyl tin |
HR-TGC1 | Phenylmethyl silicone | Low | 360 °C / 390 °C | Triglyceride |