OPERATING SPECIFICATIONS |
Number of eluents |
1515: One
1525: One or two |
Programmable flow rate range |
0.00 to 10.00 mL/min in 0.01 mL increments |
Flow precision |
≤ 0.1% RSD or 2 sec SD, six replicates, based on retention time or volumetric measures, at 1.0 mL/min, 1000 to 2000 psi backpressure, PQ test (each pump individually) |
Flow accuracy |
± 1.0% of setting at 1.0 mL/min or 30.0 μL/min, whichever is greater, using degassed methanol, 1000 to 2000 psi backpressure (each pump individually) |
Maximum operating pressure |
6000 psi (41,370 kPa, 401 bar) |
Operating pressure limits |
Software programmable high and low pressure limits; user selectable in psi, kPa, bar < |
Delay volume (1525) |
200 μL (with one GM 150 mixer) |
Gradient compositional accuracy (1525) |
± 0.5% of setting at 1 mL/min and 1000 psi backpressure (methanol:methanol with propylparaben) with one GM 150 mixer |
Gradient compositional precision (1525) |
< 0.5% RSD of setting at 1.0 mL/min and 1000 psi back pressure (methanol:methanol with propylparaben) with one GM 150 mixer (based on six replicates of compositional accuracy) |
Pressure ripple (one pump) |
≤ 2.0% at 1.0 mL/min, degassed methanol, at 1000 psi backpressure |
Gradient ripple |
< 1.5% normalized to full scale between 10% to 90%, 1.0 mL/min, and 1000 psi backpressure (methanol:methanol with propylparaben with one GM 150 mixer) |
1525μ Binary HPLC Pump |
OPERATING SPECIFICATIONS |
Programmable flow rate |
0.00 to 5.00 mL/min, in 0.01 mL/min increments |
Flow precision |
2 sec SD, six replicates based on retention time or volumetric measures at 0.2 mL/min, 1000 to 2000 psi back pressure, PQ test (each pump individually) |
Flow accuracy |
2.0% of setting at 0.1 mL/min using degassed methanol at 1000 to 2000 psi backpressure (each pump individually) |
Maximum operating pressure |
6000 psi (41,370 kPa, 401 bars) |
Operating pressure limits |
Software programmable high and low pressure limits; user selectable in psi, kPa, bar < |
Delay volume |
30 μL (without mixer), < 100 μL (with one 50 μL mixer) |
Gradient accuracy |
< 1.0% of setting (typical) at 0.1 mL/min between 10% to 90%, and 1000 psi backpressure (methanol:methanol with propylparaben and one 50.0 μL mixer) |
Gradient precision |
< 0.5% RSD (typical) between 10.0% to 90.0%, 0.1 mL/min, and 1000 psi backpressure (methanol:water with octanophenone and one 50.0 μL mixer) |
Pressure ripple (one pump) |
< 2.0% at 0.5 mL/min, degassed methanol, at 1000 to 2000 psi backpressure |
Gradient ripple |
< 1.5% normalized to full scale between 10.0% to 90.0%, 0.1 mL/min, and 1000 psi backpressure (methanol:methanol with propylparaben) with one 50.0 μL mixer |
INSTRUMENT CONTROL |
Communications |
Ethernet, IEEE-488 |
External control |
Empower™ Breeze™ or MassLynx™ Software |
Event I/O |
Back of instrument Detachable terminal strip |
PHYSICAL / ENVIRONMENTAL SPECIFICATIONS |
Physical size (without bottle holder) |
Height: 30.5 cm (12.0 in.)
Depth: 61.0 cm (24.0 in.)
Width: 43.0 cm (17.0 in.) |
Weight |
1515: 20.4 kg (45.0 lbs)
1525: 27.2 kg (60.0 lbs) |
Wetted surface material |
316 stainless steel, sapphire, reinforced fluorocarbon polymer seals, carbon-reinforced Tefzel < |
Acoustic noise |
70 dB(A) |
Operating temperature range |
4 to 40 °C |
Operating humidity range |
20% to 80%, non-condensing |
1500 Series Manual Injector (Applicable for 1515 and 1525) - Rheodyne 7725I Injector |
OPERATING SPECIFICATIONS |
Inject switch |
Contact closure, pre-wired |
Mounting |
Integral for 1500 Series HPLC pumps |
Sample holdup |
Zero |
Flow during switching |
Continuous, make-before-break |
Injection |
Partial or full loop |
Loop size |
20 μL (standard)
Changeable (5-, 50-, and 200-μL loop supplied) |
Wetted materials |
316 ss, ceramic, inert polymers |
Waters 1500 Series Column Heater |
OPERATING SPECIFICATIONS |
Set point temperature range |
20 to 60 °C; set ≤ 5 °C above ambient temperature |
Temperature accuracy |
± 0.80 °C |
Temperature precision |
± 0.25 °C |
Pre-column heating of mobile phase |
± 0.50 °C of column temperature up to 5.00 mL/min |
Pre-heat tube volume |
Approximately 35 μL |
Column capacity |
Up to four 7.8 mm x 300.0 mm without guard columns
Two columns with guard columns |
PHYSICAL SPECIFICATIONS |
Dimensions |
Height: 43.0 cm (17.0 in.)
Depth: 35.6 cm (14.0 in.)
Width: 15.2 cm (6.0 in.) |
Weight |
5.9 kg (13.0 lbs) |
ELECTRICAL SPECIFICATIONS |
Input voltage range |
120/240 VAC |
Input frequency range |
50/60 Hz |
INSTRUMENT CONTROL |
Communications |
Ethernet, IEEE - 488 |