Welcome, Sign In First time visitor? Create a New Account
Your cart is empty

Waters ACQUITY UPLC 30 cm Column Heater/Cooler (30-cm CHC) - 186015011

Price:
Live chat below or contact us for latest price.


Download Data Sheet

ACQUITY UPLC 30 cm Column Heater / Cooler (30-cm CHC)

Please contact our Sales Team for a quote.

Ease conversion to UPLC® technology in your laboratory with the ACQUITY UPLC® 30 cm Column Heater/Cooler (CHC). Run HPLC columns and methods on the ACQUITY UPLC System, and transition your HPLC temperature-controlled HPLC methods to UPLC methods, while preserving current efficiency.

The ACQUITY UPLC 30 cm CHC enables your lab to upgrade your LC instrumentation to the newest chromatographic capabilities, while retaining your current methods until you have the time and resources to fully transition to UPLC.

The design of the 30 cm Column Heater/Cooler makes it particularly convenient for users who are familiar with the popular Alliance® HPLC Column Heater/Cooler, upon which the 30 cm CHC is based.

Features:

  • Heating and cooling from 4 to 65 °C
  • Temperature control of up to four HPLC columns with lengths up to 300 mm and internal diameters up to 7.8 mm
  • Columns can be plumbed in series
  • Qualification support with Waters AQT™ and SystemsQT™ products
  • Empower™ and MassLynx™ method and instrument control
  • Customized tubing kit configured exactly for installation on the ACQUITY UPLC system
  • Can be installed on new or existing ACQUITY UPLC Systems

Specifications:

OPERATING SPECIFICATIONS
Number of columns Accommodates multiple columns, serially connected columns can also be supported (up to six)
Column sizes 20 to 300 mm length x 2.1 to 7.8 mm internal diameter (I.D.), plus either a guard column or an in-line filter (20 mm maximum)
Compartment temperature control 4.0 to 65.0 °C in 0.1 °C increments
Compartment temperature measurement accuracy ±0.8 °C (between 20.0 to 50.0 °C) ± 1.0 °C for remaining range
Note: temperature is measured by an NIST traceable probe located next to the measurement sensor will be within the specification.
Compartment operational temperature range Cooling: 15 °C below ambient
Heating: 5 °C above ambient to 65 °C
Communications Controlled by ACQUITY UPLC Sample Manager
Control software ACQUITY UPLC console software v1.30 or higher
PHYSICAL Specifications
Dimensions Width: 20.3 cm (8.0 in.) 
Height: 55.2 cm (21.8 in.)
Depth: 35.6 cm (14.0 in.) without cables
Weight 13.6 kg (30.0 pounds)
ENVIRONMENTAL Specifications
Operating temperature range 4 to 40 °C (39 to 104 °F)
Operating humidity range 20% to 50%, non-condensing
Acoustic noise < 65 dBA (instrument generated)
Airflow requirements Clearance of 15 cm (6 in.) at the module’s rear
Ordering Information
ACQUITY UPLC Column Heater/Cooler kit 186015011

 


  • Model: 186015011
  • Manufactured by: Waters