Welcome, Sign In First time visitor? Create a New Account
Your cart is empty

LNI SWISSGAS SONIMIX 3012/3022 Multigas calibrator with GPT, ozone & self re-generable zero air generator

Price:
Live chat below or contact us for latest price.


Download Data Sheet

SONIMIX 3012 / 3022 Multigas calibrator with GPT, ozone & self re-generable zero air generator

Available in the Following Configurations:

Sonimix 3022 3pts 3 points precision calibrator. 19" Rack casing (standard dilution rates)*
Sonimix 3022 10pts 10 points precision calibrator 19" Rack casing (alternative) Dilution rate from 1/100 to 1/1000*
Sonimix 3022 10pts 10 points precision calibrator 19" Rack casing (alternative) Dilution rate from 1/200 to 1/2000*
Sonimix 3012 3pts 3 points precision calibrator portable casing (standard dilution rates)*
Sonimix 3012 10pts 10 points precision calibrator portable casing (alternative) Dilution rate from 1/100 to 1/1000*
Sonimix 3012 10pts 10 points precision calibrator portable casing (alternative) Dilution rate from 1/200 to 1/2000*

*Module for 4 gases Inlets (3 additional Inlets) included

The multigas calibrator with GPT, ozone and self re-generable zero air generator, Sonimix 3012/3022, is an instrument designed to calibrate manually or by remote control up to 6 monitors installed in an air pollution monitoring station, such as SO2, HC, BTX, CO, NOx and O3.

The Sonimix 3012/3022 is equipped with:

  • an inlet for high concentration mixtures of diluted gases such as NO, NO2, SO2, CO, HC, BTX or mixtures of several compatible polluants in the same cylinder.
  • a zero air generator composed of an internal compressor with self re-generable filters that removes moisture, O3, NOx, SO2, HC, BTX and optionally CO and CH4.
  • a binary gas dilutor with two inlets (diluted gas and air), based on sonic nozzles (ISO 6145-6 standard)
  • The ozone generator is regulated by UV-light feedback and compensated from the atmospheric pressure and ambient temperature variations, with alarm on the drift of UV-light as well as the circuits for the GPT.

Product features:

  • Accuracy - (0,6% rel.)
  • Best quality price ratio
  • No maintenance nor pressurised Air cylinder with its included self re-generable zero air generator
  • Sonic nozzle technology
  • Remote control
  • Ergonomic with LCD display

Main applications:

  • Air pollution monitoring

Voltage settings:

  • 6700 13 867 - Voltage 230V/50-60 Hz
  • 6700 13 868 - Voltage 115V/60 Hz
  • 6700 13 869 - Voltage 100V/60 Hz (not available with catalysis oven)

Options:

  • 6800 12 030 - Ozone generator from 50 to 500ppb and circuit for the GPT
  • 6800 08 015 - Integrated self regenerable Zero Air generator
  • 6800 08 017 - Catalyser oven for CO and CH4
  • 6800 10 004 - Cold filter for CO (no possible if catalytic oven)
  • 6800 10 000 - Portable casing for rack 19''
  • 6980 40 040 - ISO 17025 New device Additional price for ISO 17025 Calibration (10 points)

Consumables:

  • 6700 13 870.6 - Set of the consumables filters for the annual maintenance Sx 3022 (2xAlumine; 2 Pura/AC)
  • 6700 13 870.7 - Set of the consumables filters for the annual maintenance Sx 3012 (2x Alumine/Pura/AC)
  • 3741 40 020 - Dust pump filter
  • 3741 49 001 - Filter for the automatic purge V5
  • 3760 00 902 - Kit of diaphragm and valves (popets) for the compressor
  • 6704 06 176 - Refill for CO filter

Technical Specifications:

Models: Sonimix 3012/3022 Sx 3012/3022 Sx 3012/3022
010
General data
Dilution point 3 10
Dilution range 1/166, 1/250, 1/500
(100, 200, 300 ppb with 50 ppm at
inlet)
1/200 to 1/2000 or 1/100 to 1/1000
Dilution principle Sonic nozzle
Dilution gas line flow accuracy 0,6% relative
Repeatability on flow Better than 0.3% relative
Warm up time < 45 min
Stabilization time 3 min
Operating temperature 15°C - 40°C
Dimensions 19" 3HE/84 TE , 500mm deepn portable casing in option
Net weight 20 Kg
Gas inlet requirments
Zero Air Requires high grade zero air from internal zero air generator or from cylinder DP < -40°C
Carrier gas NO, NO2, O3, SO2 < 1ppb
BTEX < 1 ppb
CH4 < 50 ppb & CO
HCnm < 50 ppb if catalysis oven is installed
Dilution gas Requires High concentrated premixtures
SO2 in Air - NO in N2 - CO in Air - HC or BTX in Air
Pressure 3 ± 0.3 bar rel
Fittings 1/4’’ and 1/8’’ Swagelok
Gas outlet requirments
Flow About 3000 Nml/min, other values upon request
Fittings 1/4’’ and 1/8’’ Swagelok
O3 section
O3 concentration 25 - 500 ppb, adjustable by software
Accuracy Better than +/- 2 ppb from 25 to 200 ppb
Repeatability on flow Better than +/- 2.5 ppb (at 250 ppb O3)
Communication
Key board and VFD display Standard
RS232 Standard
Electrical data
Power supply voltage 230V/50-60Hz
115V/60Hz
100V/50-60Hz
Installed power (Max) 65 - 150W

  • Model: SX 3012/3022