The photometric accuracy of microplate readers can be checked using the calibration standard 666R013. The microplate reader standard has the same dimensions as a microplate and has windows with a 6.6 mm diameter at the same positions as the 96 wells. Neutral density glass strips with varying absorption values are located behind groups of 16 windows. All calibration standards come with a certificate of calibration.
666R013 Microplate reader calibration standard
The photometric accuracy of microplate readers can be checked using the calibration standard 666R013. The microplate reader standard has the same dimensions as a microplate and has windows with a 6.6 mm diameter at the same positions as the 96 wells. Neutral density glass strips with varying absorption values are located behind groups of 16 windows. All calibration standards come with a certificate of calibration.
Type | Description | Nominal Absorbance | Absorbance at nm |
666R013 | Microplate Reader Calibration Standard, NG 11, NG 5, NG 4, NG 3 Certified | 0,25; 0,5; 1,0; 1,5; 2,5 | 405, 450, 490, 650 |
For checking the wavelength accuracy of microplate readers microplate reader standards are available, where a neutral density glass stripe is substituted by a homium oxid glass stripe (on customer request).
- Model: 666R013
- Manufactured by: Hellma