Welcome, Sign In First time visitor? Create a New Account
Your cart is empty

Hellma Potassium dichromate in Perchloric Acid (430 nm)

Price:
Live chat below or contact us for latest price.


Potassium dichromate in Perchloric Acid (430 nm)

 

Description Potassium dichromate filter, liquid filter, certified
Use Assessment of photometric accuracy (absorbance) in the Vis range (measuring wavelength 430 nm) with a spectral bandwidth of 2 nm or less
Material 0.06% potassium dichromate dissolved in perchloric acid and sealed in Hellma precision cells made from quartz SUPRASIL®

 

667-UV600 0.06% Potassium dichromate dissolved in perchloric acid

Fifth revision of European Pharmacopeia stipulates also checking the photometrical accuracy of the spectrophotometer at a wavelength in the visible region (430 nm) besides the checking in the UV region. The testing has to be carried out by means of a solution of 0.06% potassium dichromate dissolved in perchloric acid. The absorbance value of the filter is determined at a wavelength of 430 nm and is specified in the certificate. It should be noted that the measurement is carried out with the filter 667-UV14 (0.01 N perchloric acid) in the reference beam path. Measured values of the reference filter are separately displayed.

 
Typical spectrum of a 0.06 % potassium dichromate solution
 

 

Wavelength
(nm)
Specific Absorption
1 %1 cm
Maximum Tolerance
1 %1 cm
430 15,9 15,7 to 16,1

  • Model: 667-UV600
  • Manufactured by: Hellma