Hitachi Chromaster 5110 HPLC Pump
Chromaster HPLC 5110 pump: Pressure up to 40 MPa
Please contact our Sales Team with your laboratory requirements and specifications including PC equipment and Software. Uvison service contracts are also available - please see our Engineering section for further information -https://uvison.com/engineering |
This pump is available in the following configurations - please specify when ordering:
- Chromaster HPLC 5110 pump with manual purge valve. 400 bar max. operating pressure (9,99 ml/min max. flow rate)
- Chromaster HPLC 5110 pump with auto-purge valve. 400 bar max. operating pressure (9,99 ml/min max. flow rate)
Excellent solvent delivery performance
One of the most important performance measures for HPLC is retention time reproducibility. Excellent gradient performance resulting from the highly accurate solvent delivery by the 5110/5160 pump with Hitachi’s unique high-speed feedback real-time control system and High Frequency Mode (HFM) of the proportioning valve make the high retention time reproducibility possible.
Pump Optional Accessories Available - please specify when ordering:
- Low-pressure Gradient Unit for 5110/5160 (with Conventional Mixer)
- 6-channel Degassing unit (480 μl / ch)
- Plunger Washing Pump
- Conventional Mixer (700 μl)
- Semi-micro Mixer (200 μl)
- Dynamic Mixer (2,000 μl)
- Manual Injector Holder
- Column Holder
- UI Pad for 5110/5160
- AC adapter (150 W)
- Hexane 6-channel Degassing unit for 5110/5160
- Hexane Check Valve Set for 5110
- THF kit for 5110/5160 - Withstand pressure 40 MPa
- THF kit for 5110/5160 with AP valve - Withstand pressure 40 MPa
- THF kit for Low Gr unit for 5110/5160
SPECIFICATIONS:
Pump CM 5110 / CM5110 | |
Maximum operating pressure | 40,0 MPa (0,001 – 5000 ml/min); 20,0 MPa (5001 – 9999 ml/min) |
Flow rate range | 0,001 to 9999 ml/min |
Flow rate accuracy | ±2 µl/min (0,01 - 0,1 ml/min); ±1% (0,101 - 5 ml/min) |
Flow rate precision | 0,075% RSD |
Optional accessories | 6-channel degassing unit; 4-channel low pressure gradient; additional internal plunger washing pump; UI-PAD (Keyboard) |
Pumping system | Dual plunger reciprocating pump system Series connection, pulsation eliminatuon system |
Operating flow late range | 0.001 to 9.999 mL/min |
Maxmum operating pressure | 40 MPa (0.001 to 5.000 mL/min) 20 MPa (5.001 to 9.999 mL/min) |
Flow rate accuracy | ±1.0% or ±2.0 μL /min, whichever is greater (0.010 – 5.000 mL/min, under a specified condition) |
Flow rate precision | SD0.02 min or RSD0.075%, whichever is greater, under a specified condition |
Materials of wetted parts | SUS316, ruby, sapphire, ceramics, PTFE, carbon-contaning PTFE, PEEK (Auto-purge valve unit) |
Functions of GLP |
|
Dimentions and weight | 340 (W) × 440 (D) × 140 (H) mm, Approx.16 kg |
Power supply and Power consumption | DC 24 V, 4 A (Maximun) 96 W (power supply from organizer) |
Ohters | Pumps are available with and without an auto-purge valve. |
Also Available:
Note
- All pumps include leak sensor
- Need to configurate degasser and low pressure gradient
- 400 bar spare parts available
- Model: CM 5110 / CM5110
- Manufactured by: Merck Hitachi