Welcome, Sign In First time visitor? Create a New Account
Your cart is empty

Hitachi Chromaster 5210 HPLC Autosampler

Price:
Live chat below or contact us for latest price.


Download Data Sheet

Hitachi Chromaster 5210 HPLC Autosampler

Chromaster 5210 Autosampler

Please contact our Sales Team with your laboratory requirements and specifications including PC equipment and Software.

Uvison service contracts are also available - please see our Engineering section for further information - https://uvison.com/engineering 

This autosampler is available in the following configurations - please specify when ordering

  1. 5210 Autosampler without Thermostat
  2. 5210 Autosampler with Thermostat

Extremely low carry-over

The first hurdle to be overcome in reducing the amount of carry-over is to create a structure by eliminating the dead volume in the autosampler flow path. 5210 autosampler represents a revamping of the basic structure to minimize the dead volume. Additionally, active wash of the needle outer wall by dedicated pump provides constant washing effect. The result is extremely low carry-over.

Excellent injection volume reproducibility supporting reliable analysis The syringe and syringe moving part are optimized for each of the loop injection and direct injection systems. As a result, the measurement accuracy by the syringe is improved, resulting in excellent injection volume reproducibility

Easily accommodates a 300 mm analytical column fitted with a pre-column

The door, which opens in an L-shape pattern and with internal dimensions 375 mm wide and 114 mm high, facilitates the connection and stowing tasks for pre-column equipped column. The oven can accommodate up to three 300 mm columns. The column installation space, which has an air circulation system, permits easy mounting and detaching of columns.

Pre-heating function and wide temperature control range

The block-type pre-heating function based on Peltier heating and cooling control, delivers excellent peak symmetry and shape.*1 Also, the oven with a capability to regulate*2 temperature from 15 degree below ambient temperature to ambient temperature +60? can accommodate various applications.

ADDITIONAL ACCESSORIES

  1. Hexane kit for 5210/5260
  2. Chromaster HPLC THF resistant autosampler kit for 5210/5260 - Withstand pressure 40 MPa
  3. Interface control board (IFC board) (for installing a 5210 autosampler)
  4. Sample rack (4 mL × 72)
  5. Thermostat rack (4 mL × 72)
  6. Sample rack (1 mL × 195)
  7. Thermostat rack (1 mL × 195)
  8. Micro plate rack (2 pcs)
  9. Thermostat micro plate rack (2 pcs)
  10. Syringe kit (70 μL)
  11. Sample loop kit (5 μL)
  12. Sample loop kit (10 μL)
  13. Sample loop kit (20 μL)
  14. 2-channel Degassing unit (250 μL / ch)
  15. AC adapter (150 W)
  16. UI Pad for 5210

SPECIFICATIONS

Sample capacity 195 × 1 mL
120 × 1.5 mL
72 × 4 mL
2 × MTP (96,384)
Sample injection system Loop injection metyhod
(Cut injection, All volume injection, Full loop injection method)
Syringe volume 175 μL (standard) (There is an optional syringe.)
Sample Injection volume 0.1?50 μL (100 μL loop) (standard)
0.1?100 μL ( 200 μL loop) (accessory of 5210 Autosampler)
Injection volume precision 0.2%RSD (10 μL, cut injection method )
0.25%RSD (5 μL, cut injection method)
0.9%RSD (1 μL, cut injection method)
1.0%RSD (1 μL, All volume injection method)
????.2%RSD (5 μL, full loop method)
Carryover 0.003 (cut method)
Materials of wetted parts SUS316, Vespel® (Polyimide resin), fluororesin,
PP, EPDM, FPM, perfluoroelastomers
Withstand pressure 40 MPa
Temparature setting range 1 to 45? (1? step), using 5210 Autosampler with a thermostat
Temparature controll range [RT-21?] to [RT+25?] and range of the temparature setting (with a vial)
[RT-15?] to [RT+20?] and range of the temparature setting (with a MTP)
(using 5210 Autosampler with thermostat)
Functions of GLP (a)Injection port seal (b)Injection valve seal (c)Syringe valve seal
(d)Syringe (e) Wash pump operation time
Dimentions and weight 340 (W) × 440 (D) × 280 (H) mm, Approx.24 kg
(with thermostat, approx.29 kg)
Power supply and Power consumption DC24 V, 4 A (Maximum)/96 W (power supply from organizer)
AC100 to 240 V (50 Hz/60 Hz)
110 VA (using 5210 Autosampler with thermostat)
Others Autosamplers are available with and without a thermostat.

 


  • Model: CM 5210 / CM5210
  • Manufactured by: Merck Hitachi