Hitachi Chromaster 5610 Mass Selective MS Detector

Price:
Live chat below or contact us for latest price.


Download Data Sheet

Hitachi Chromaster 5610 MS Detector

Chromaster 5610 mass selective detector, including control software (MSD System Manager, English Version)

Please contact our Sales Team with your laboratory requirements and specifications including PC equipment and Software.

Uvison service contracts are also available - please see our Engineering section for further information - https://uvison.com/engineering 

The Chromaster 5610 MS Detector is a new quadrupole mass spectrometry-based detector which offers greater accuracy, convenience, and compact size for HPLC users. The reliability of qualitative analysis is increased significantly by mass information via MS detector. The system offers many advantages even for first time HPLC users such as the same footprint as Chromaster series, low N2 gas consumption, and additionally provides greatly improved maintainability by using a newly developed atmospheric pressure ion filter.

FEATURES

Optimized for qualitative analysis
When measuring samples that do not absorb UV light or analyzing target compounds that cannot be identified by using UV spectra alone, additional information provided by mass spectra can improve the reliability of qualitative analysis.

Ease of use
Operability on par with standard HPLC systems. The atmospheric-pressure ion filter allows for replacement or cleaning of the filter without stopping the vacuum pump during maintenance.

Compact design for small footprint
The space-saving footprint is equivalent to that of a standard HPLC system. The required power source is 200 to 240 VAC,
allowing more flexibility when choosing the installation site. As the system is designed to use a minimum amount of N2 gas (maximum flow rate of 3,4 l/min), even an N2 gas cylinder can provide sufficient gas.

OPTIONAL ACCESSORIES

  1. AC adapter (150 W)
  2. Integrated module control
  3. UI Pad
  4. System piping kit

SPECIFICATIONS

5610 MS Detector
Measurement mass range (m/z) 20 to 1,000
Ion source Electrospray ionization (ESI)
Dimensions and Weight 440(W) x 610(D) x 430(H) mm / approx. 51 kg
Power supply and Power consumption AC 200 to 240 V (50 Hz/60 Hz)/1,000 VA
N2 gas usage Max flow rate 3,4 l/min, pressure 300 ± 20 kPa

CHROMASTER DETECTORS AVAILABLE 


  • Model: CM 5610 MS / CM5610
  • Manufactured by: Merck Hitachi