High-Performance Liquid Chromatograph Primaide
Primaide PM1410 HPLC Variable Wavelength UV Detector with Autocalibration
Please contact our Sales Team with your laboratory requirements and specifications including PC equipment and Software. Uvison service contracts are also available - please see our Engineering section for further information -https://uvison.com/engineering |
The optics of the Primaide 1410 UV detector allows data collection with lower noise and drift. The lamp and flow cell are situated in a front-access layout and are easily replaceable.
Low noise for high-sensitivity detection
The Primaide 1410 UV detector achieves a noise level of no more than 0.6 × 10 -5 AU, greatly enhancing trace component analysis.
High response and low noise for rapid analysis
- Low-noise measurements in 0.05 seconds.
- High-sensitivity detection rapidly yielding sharp and accurate peaks.
Stable, without influence from ambient temperature changes
Since local environments may vary considerably, the optics of the UV detector is well designed to avoid the effects of daily and seasonal temperature changes, ensuring stable and reproducible results.
Wavelength measurement in the UV region with a built-in Hg lamp
Use the 254-nm emission line of the mounted Hg lamp to test the accuracy of wavelength measurement in the UV region, which is the most utilized region in liquid chromatography. This test is conducted using 3 wavelengths from the Hg and D2 lamps, enabling highly reliable analyses across a broad wavelength range.
Front access layout and easy maintenance
Both the lamp and flow cell are replaced through the front of the module, allowing easy maintenance, regardless of where the instrument is installed.(Lamp replacement)
A DAD for the Primaide is also available.
SPECIFICATIONS
Primaide1410 UV | |
Optical system | Double-beam ratio photometry system |
Light source | D2 lamp. Hg lamp (for wavelength check) |
Wavelength range | 190 to 600 nm |
Wavelength accuracy | ± 1 nm |
Noise | 0.6 × 10-5 AU or less (250 nm) |
Drift | 1.0 × 10-4 AU/hr or less (250 nm) |
Response | 0.05, 0.1, 0.5, 1, 2, 4, 8 sec |
Dimensions/weight | 340(W) × 464(D) × 159(H) mm, approx. 12.6 kg |
- Model: 1410
- Manufactured by: Merck Hitachi