High-Performance Liquid Chromatograph Primaide
Primaide™ PM1430 diode array detector (DAD)
Please contact our Sales Team with your laboratory requirements and specifications including PC equipment and Software. Uvison service contracts are also available - please see our Engineering section for further information - https://uvison.com/engineering |
Photo Diode Array Detector System
The Primaide 1430 Diode Array Detector is as sensitive as a UV detector. The DAD system has many functions, including spectrum collection, chromatography using multiple wavelengths, a peak purity test, a spectral library, and more.
High sensitivity
Novel optics and improved software reduce noise considerably (0.5 × 10-5 AU or less). The reputation, "DADs are less sensitive," is now obsolete.
Minimizing the influence of temperature changes in the measurement cell
The light source is stabilized via a large lamp housing. The spectrometer and housing are separated and thermally insulated. By cooling the lamp housing separately, the influence of ambient temperature is minimized. The new design incorporates temperature sensors as well as an associated variable rotation speed cooling fan, which largely eliminate temperature changes inside the instrument. This reduces noise and enhances stability, allowing for stable analysis day after day.
High resolution and accurate spectra
High wavelength resolution (0.78 nm/data point) is achieved with an array of 1,024 photo diodes. Uniform resolution is maintained throughout a wide range of wavelengths from the UV to VIS, thanks to the diffraction optics, which features a grating.
Digital noise reduction in rapid analysis
Data with a low level of noise are obtained in a very short sampling period (0.05 seconds). Noise-free, sharp peaks are obtained during rapid analysis.
Automatic wavelength calibration
A built-in Hg lamp is utilized in the automatic wavelength accuracy test for the commonly-used UV region.
SPECIFICATIONS
Primaide1430 DAD Diode Array Detector | |
Number of photodiode bits | 1024 |
Light source | D2 lamp, W lamp, Hg lamp (for wavelength check) |
Wavelength range | 190 to 900 nm |
Wavelength accuracy | ± 1 nm |
Noise | 0.5 × 10-5 AU or less (250 / 600 nm) |
Drift | 0.5 × 10-3 AU/hr or less (250 / 600 nm) |
Spectrum acquisition cycle | Selectable from 50, 100, 200, 400, 800, 1600, 3200 msec |
Dimensions/weight | 340(W) × 464(D) × 159(H) mm, approx. 13.5 kg |
- Model: 1430
- Manufactured by: Merck Hitachi