Welcome, Sign In First time visitor? Create a New Account
Your cart is empty

Hitachi Chromaster 5310 HPLC Column Oven

Price:
Live chat below or contact us for latest price.


Download Data Sheet

Hitachi Chromaster 5310 HPLC Column Oven

Chromaster HPLC 5310 Column Oven Only - Horizontal type

Please contact our Sales Team with your laboratory requirements and specifications including PC equipment and Software.

Uvison service contracts are also available - please see our Engineering section for further information - https://uvison.com/engineering 

This column oven is available in the following configurations - please specify when ordering:

  1. Chromaster HPLC 5310 column oven for max. 3×300 mm columns
  2. Chromaster HPLC 5310 column oven with 0,5 m e-Line cable, power cable and integrated 600 bar VICI switching valve

Pre-heating function and a wide temperature control range
The block-type pre-heating function based on Peltier heating and cooling control delivers excellent symmetric and sharp peak shape. The oven can regulate*1 temperature from 15 degree below ambient temperature to ambient temperature +60 °C, and can accommodate various columns.
*1 Temperature setting range: 1 to 85 °C.

Easily accommodates a 300 mm analytical column fitted with a guard?column
The door, which opens in an L-shape pattern and with internal dimensions 375 mm wide and 114 mm high, facilitates the connection and stowing tasks for a guard-column and column. The oven can accommodate up to three 300 mm columns.

Column management system (5310 optional)
Hitachi column management system can manage the Log information on analytical columns and guard-columns from any manufacturer. Log information can be written and read through a connector mounted on the column oven or USB port in the computer. ID Tags can be used repeatedly.*3
*3 Approximate read/write life time: 100,000 times

Low volume pre-heating to suppress peak diffusion
The newly designed low volume pre-heating tube minimizes the peak diffusion while the temperature stability is maintained, resulting in high resolution analyses and high reproducibility.

  • pre-heat piping volume - 5310 39 μl / 6310 1 μl

Column Oven Optional Accessories Available - please specify when ordering:

  1. Column management system for 5310
  2. 2-position, 6-port valve for 5310
  3. 3 column selector valve for 5310
  4. UI Pad for 5310

SPECIFICATIONS:

Temperature Setting Range 1 to 85 °C (in increments of 1 °C)
Temperature Setting Accuracy ±1,0 °C (20 to 85°C in preheating)
Temperature Control Precision SD ≤0,2 °C
Temparature controll range [RT-15?] to [RT+65?] and range of the temparature setting
Columns accommodated 300 mm columns x 3 pcs
Available valves 3-column selector valve (max. 350 bar); 6-way 2-position valve (max. 350 bar); 6-way 2-position vici-valve (max. 600 bar)
Temparature controll system Heating/Cooling block?air circulation system
Functions of GLP Recording of the changeover number of times and exchange dates of the optional changeover valve.
Column capacity 300 mm × 3 (Maximun)
Dimentions and weight 410 (W) × 440 (D) ×140 (H) mm, Approx 13 kg
Time program functions ?Temparature setting ?Switchng valve (changing of position)
Power supply and Power consumption AC100 to 240 V (50 Hz/60 Hz)/230 VA (with optional valves) ?The Organizer and the AC adaptor are not necessary
Optional accessories UI-PAD (Keyboard)

Also Available:


  • Model: CM 5310 / CM5310
  • Manufactured by: Merck Hitachi