515 Isocratic HPLC Pump
Please contact our Sales Team for a quote.
The Waters 515 HPLC Pump combines the most important aspects of HPLC performance - accuracy, precision, reliability, pulse-free eluent delivery and proven functionality. An easy- to-read display and keypad user interface enable error-free programming. Automatic purge, flow calibration and on-board diagnostics enhance confidence in results.
The 515 HPLC pump is a programmable, free-standing HPLC pump available in isocratic, gradient, and semi-preparative configurations:
- Precise stepper motor control (48 steps/µL resolution) of dual reciprocating pistons
- Non-circular drive gears provide smooth, uniform eluent delivery
- User-programmable flow rate calibration and compressibility compensation factors ensure straightforward pump operation qualification
- Tests to retention time variations of less than 0.1% for highly reproducible peak performance
- Easy-to-read display and keypad for error-free programming
- Automatic purging and flow calibration
Specifications:
Operating Specifications | |
Number of Eluents | One |
Programmable Flow Rate Range | 0.000 - 10.000 mL/min., in 0.001 mL/min increments, standard 100 μL heads |
Flow Precision | ≤ 0.1% RSD or two seconds SD, six replicates, based on retention time or volumetric measures, at 1 mL/min, 1,000 - 2,000 psi backpressure, PQ test |
Flow Accuracy | ± 1.0% or 30 μL/min, whichever is greater, at 1 mL/min, degassed methanol, 1,000 - 2,000 psi backpressure |
Compressibility Compensation | Programmable |
Flow Calibration | Programmable |
Purge Function | Preprogrammed, user initiated |
Maximum Operating Pressure | 6,000 psi (41,370 kPa, 401 bar) |
Operating Pressure Limits | Programmable high and low pressure limits User selectable in psi, kPa, bar |
Pressure Ripple | ≤ 2.0% (1 mL/min, degassed methanol, 1,000 psi backpressure) |
Plunger Guiding Syste | Floating, self-aligning mount |
Primary Wetted Surface Materials | 316 stainless steel, sapphire, reinforced fluorocarbon polymer seals |
System Physical Specifications | |
Physical Size (W x H x D) | 7.5” (19.1 cm) x 8.0” (20.3 cm) x 18.5” (47 cm) |
Weight | 33 lb (15 kg) |
Environmental | |
Acoustic Noise | <70 dB(A) at operator position |
Operating Temperature Range | 4? C to 40? C |
Operating Humidity Range | 20% to 80%, noncondensing |
Power Requirements | |
Voltage Range | 85 to 264 VAC |
Frequency | 47 to 63 Hz, single phase |
Instrument Control and Communications | |
Operating Pressure Output | 0 to -10V, analog to external control device |
Pressure Limit Shutdown Output | +15V to -15V, analog to external control device |
External Control Device Input | TTL-level pulse stream from external control device |
Diagnostics | Pump use metrics retained automatically, user accessible through LCD display Plunger strokes on unit, plunger strokes on seals |
Options | |
External System Control | Waters or Micromass® software enables isocratic, flow programming, and multipump gradient modes |
Plunger Seal Wash | Continuous |
225 μL displacement volume heads | 0.000 - 22.500 mL/min, in 0.001 mL/min increments, 3,000 psi maximum operating pressure, enables higher flow rates for larger scale separations |
Ordering Information | |
Description | Part Number |
Waters 515 HPLC Pump | WAT207000 |
Waters 515 Extended Flow Conversion Kit | WAT207119 |
Plunger Wash Kit, 100 μL heads | WAT031870 |
Plunger Wash Kit, 225 μL heads | WAT030852 |
Stainless Steel Tee Fitting | WAT075215 |
Static Mixer | WAT051518 |
Reference Valve and Mounting Kit | WAT207144 |
- Model: WAT207000
- Manufactured by: Waters